Temperature And Humidity Controlled Cabinets (dehumidification) ULB-E01
- Ulabester
- China
- About 30 days
- 500 sets
Complete stainless steel material, French imported brand of compressors.
Touch screen, South Korean import, temperature humidity controller
System capacity:120 groups program X100 sections, required segmentation sections of each group can be arbitrarily divided and each program can freely link another.
Cycle setting: each operation procedure can carry out 9999 times or heaps of times cycle and repeated performance, and to the degree, divide further 5 segmentation sections to do part cycle of the extra.
Offer CE and ISO 17025 calibration certicate for FREE
Temperature And Humidity Controlled Cabinets Can custom-made according different request
Temperature And Humidity Controlled Cabinets ULB-E01
Ulabester R&D the high-tech and high accuracy Microprocessor controlled Humidity & Temperature Control Cabinet (with Dehumidifier) convention type.
Specially designed for carrying out different quality control test under controlled conditions of temperature and humidity for motors, various electrical fittings, electronic circuits and components, cloth and paper etc. & in various research labs.
Test Space Options
Capacity | 80L | 150L | 225L |
Inside dimension WxHxD(cm) | 40x50x40 | 50x60x50 | 50x75x60 |
Outside dimension WxHxD(cm) | 90x150x110 | 110X170X180 | 110x185X190 |
Key Specification
Temperature range (optional, can be custom-made) | -20~150℃ | -40~150℃ | -60~150℃ | ||
Use drier-filter to dehumidify | Capacity≥80L | Capacity ≥80L | Capacity ≥80L | ||
Use dehumidifier to dehumidify | Capacity ≥150L | Capacity ≥150L | Capacity ≥150L | ||
Additional Function | 10%~98%, can be set. (can reach 10% RH in 10~30℃) | ||||
Rate of temperature increasing | Room temperature~150℃ | About 45 minutes(non-linear no-load, about 3.5℃/minute | |||
Room temperature~-20℃ | About 45 minutes(non-linear no-load, about 1.2℃/minute) | ||||
Machine precision | Control accuracy (test precision) | Temperature | +/-0.5℃ | ||
Humidity | +/-1%RH | ||||
Distribution accuracy (uniformity) | Temperature | +/-1.0℃ | |||
Humidity | +/-2%RH | ||||
Environment temperature | 5℃-28℃ | ||||
Environment relative humidity | ≤85%RH | ||||
Power switch | Air-break switch(40A, knife switch or outlet is prohibited to use.) | ||||
Humidification water | Resistivity(≥500MΩ) | ||||
Requirement of ambient air quality | Do not contain high concentration dust and flammable gas, without strong electromagnetic radiation source nearby. | ||||
Power supply | AC1¢ 220V 50/60HZ, voltage change allowed value: ±10% of fixed voltage. | ||||
NOTE:10% RH in 10~30℃ | Capacity of 80L: drier-filter to dehumidify, the test can sustain1-3 day Capacity of 150-225L: dehumidifier to dehumidify with better effect, and the test time can be 3-7days |